Selamat Datang

Ada yang bisa di bantu?

New Products > MEK > Transportable Ventilator > MTV1000 Pneuma2

MTV1000 Pneuma2


Call

Alat Medik menjual MTV1000 Pneuma2. MTV1000 Pneuma2 adalah Tipe Transportable Ventilator yang memiliki fitur Various Mode ( Dual Limb, Single Limb ), Various Patient Circuit ( Single Limb Circuit, Dual Limb Circuit ), Transportable Function ( Low Flow O2 & High Pressure O2 ), High Efficient ( Peak Inspiratory pressure up to 80cmH2O (above PEEP), Peak Inspiration Flow up to 160LPM, FiO2 monitoring (real time monitoring), Internal Air/O2 Mixer (O2 Blender), Hemo- Dynamic functions with optional Spo2 or EtCO2 functions ) dengan beberapa manfaat salah satunya dengan option ( Hemodynamics with SpO2, Nasal Cannular for O2 Stream, Proximal Sensor ). 

Tipe ini berteknologi ACPAP+ (Adaptive CPAP Plus Pressure Support)®, ACPAP+ (Adaptive CPAP Plus Pressure Support)®, ABiPAP + (Adaptive BiPAP Plus Pressure Support), Benefit of EtCO2 option in Ventilator (Hemo-Dynamics) , Benefit of SpO2 Option in entilator, PRVC (Pressure Regulated Volume Ventilation), O2 Stream, dan Proximal Flow Sensor in Pneuma.

 

 Untuk lebih jelasnya anda bisa melihat spesifikasinya dibawah ini.

1. Short Form Specifications Transport Ventilator [ER, ICU(Child to Adult)]
        7" Color TFT, Touch Screen & Knob
        O2 Blender
    Mode  Dual Limb PACV, PSIMV, VACV, VSIMV, Spont, Auto, O2 Stream®, CPAP,
        Apnea Back-up ventilation
      Single Limb ACPAP+, BiPAP+, O2 Stream®
      Optional O2Stream® (Nasal Cannular), SpO2, EtCO2
        * Additional Battery : Over 6 hours
2. Detail Specifications
  2.1 Ventilator Data (Setting Parameters)
    Predicted Body weight   10 to 150kg (Volume)
        3 to 150kg (Pressure)
    Tidal Volume   100 to 2500 ml
    Inspiratory pressure   5 to 80 cmH2O
    Pressure support   0 to 60 cmH2O
    Respiratory rate   2 to 60 bpm
    Inspiratory time   0.3 to 4 sec
    PEEP/CPAP  

0 to 45 cmH2O

    Enable inhalation trigger (En-sense) 10 to 80% of set tidal volume
    Exhalation trigger Sensitivity (Ex-trig) 10 to 99% of exhaltion time
    Trigger sensitivity   Pressure : OFF, 0.5 to 20 cmH2O
        Flow : OFF, 0.5 to 20 lpm
    FiO2%   21% to 100%
    BTPS   OFF, MANUAL DRY, MANUAL HUMIDI
    MASK   ON / OFF
    MASK Type   Full Face, Nasal, Pillow, User in ACPAP+ and BiPAP+
    ACPAP+ Mode    
    PAP   3.0 to 20.0 cmH2O
    PS   0 to 60 cmH2O (PAP+PS < 80CmH2O depend on leakage flow)
    Exsense   1 to 80%
    Brk   0.1 to 2.0 sec
    Ramp time   OFF, 1 to 60 min
    Wake up   ON / OFF
    ABIPAP+ Mode    
    I-PAP   3 to 45 cmH2O
    E-PAP   0 to 42 cmH2O
    Rate   2 to 30 bpm
    I-TIME   0.3 to 24.0 sec
    PS   0 to 45 cmH2O
    Exsense   1 to 80%
    Brk time   0.1 to 2.0 sec
    Trigger sensitivity Pressure OFF, 0.1 to 20 cmH2O
      Flow OFF, 0.1 to 20 lpm
    O2 Stream Mode    
    Flow   5 to 50 lpm
    Inspiratory Pressure limit   5 to 80 cmH2O in O2 Stream
    Trigger sensitivity Pressure OFF, 0.1 to 5.0 cmH2O
      Flow OFF, 0.1 to 5.0 lpm
  2.2 Display Data
    Parameters   Setting parameters/patient status parameters/Alarm status/I:E ratio
    Graphic Waveform  

Pressure-Time, Flow-Time, Volume-Time

    Alarm   Airway pressure, Tidal volume, Minute volume, Respiratory rate
        Apnea, O2%, Airway leak, Circuit open, Battery indicator
        O2 supply pressure failure, Ventilation in-operation
  2.3 Alarm Setting Parameters (MAIN)
    High tidal volume (Expiratory volume) OFF, 20 to 2500 ml
    Low tidal volume (Expiratory volume) OFF, 10 to 2400 ml
    High minute volume 0.5 to 50 lpm
    (Expiratory minute volume)  
    Low minute volume 0.0 to 49 lpm
    (Expiratory minute volume)  
    High respiration rate   2  to 150 bpm
    Low respiration rate   0  to 148 bpm
    High peak airway pressure   OFF, 1 to 120 cmH20
    Low peak airway pressure   OFF, 0 to 110 cmH20
    High O2%   OFF, 0 to 100%
    Low O2%   OFF, 0 to 80%
    Airway leak   OFF, 10 to 500 ml
    Apnea   OFF, 10 to 60 sec.
    O2 supply pressure Fail    
    Ventilator in-operation    
    (Option, SPO2)        
    High SpO2   OFF, 52 to 100%
    Low SpO2   OFF, 51 to 100%
    High pulse rate (SpO2)   30 to 250 bpm
    Low pulse rate (SpO2)   25 to 245 bpm
    (Option, EtCO2)    
    High EtCO2   0.0 to 10.0 mmHg
    Low EtCO2   OFF, 0.0 to 9.9 mmHg
    High inspired CO2   0.0 to 10.0 mmHg
    Low inspired CO2   OFF, 0.0 to 9.9 mmHg
    High respiration rate (EtCO2)   3 to 150 bpm
    Low respiration rate (EtCO2)   2 to 149 bpm
  2.4 Key, Indicators        
    Inspiration hold   Measures patient's lung compliance and resistance
        Elasticity, Time constant
    Key lock   Key & LCD Touch Screen Lock
    Graph freeze        
    Alarm silence  

Mutes audible alarms for 2 minutes

    Alarm reset   Clears visual indicators and messages
  2.5 Electrical Requirements        
    Power   AC : 100-230 VAC, 50/60 Hz (Free voltage)
        DC : 10.8V 8.8A, Internal battery
            Provides Max. 180 minutes of normal operation
  2.6 Physical Characteristic        
    Weight   8 kg (main unit)
    Dimension   25 x 15.6 x 37.1 (Width x Depth x Height) cm
  2.7 Environmental Requirements        
    Atmospheric condition   700 to 1060 mbar (10.2 ~ 15.4 psi)
    Altitude   -443 ~ 3048 m (-1350 ~ 10,000ft)
    Height   Up to 6560 m (20,000 ft)
    Conditions of use   10 to 40 ℃ (50 ~ 104 ℉)
    Relative humidity   10 to 85%
  2.8 Pneumatic Requirements        
    Oxygen/Air inlet supply pressure   2.46 to 6.32 kgf/cm2 / 2.4 to 6.2 bar (35 to 90 psig)

PERMANA PUTRA MANDIRI
Jl. Taman Mini Indonesia Indah (TMII) Pintu II atas No. 9. Cipayung, Jakarta Timur - Indonesia
Fax : +62-021 840 7865. Phone (Office Hours) : (+62-21) 841 5976. Cellular : +62-812 8097502, (Voice)