New Products > MEK > ICU Ventilator > MV2000 SU:M3 ( Unit Only )

MV2000 SU:M3 ( Unit Only )


Rp. 750.000.000,-

Alat Medik menjual ICU VENTILATOR type MV2000 SU:M3. MV2000 SU:M3 mempunyai spesifikasi yang hampir sama dengan type SU:M2 seperti EtCO2 ( Hemo-Dynamics ), Sp02, O2 Stream, dan Proximal Flow Sensor didalam ventilator namun ditype ini ada 2 teknologi baru, seperti Auto Vent dan PRVC atau bisa dibilang Tekanan Diatur Volume Ventilator. Untuk lebih jelasnya anda bisa melihat spesifikasinya dibawah ini.

 

 

 

 

 

 

 

 

1.

Short Form Specifications ICU Ventilator (Infant to Adult)
        12.1" Color TFT, Touch Screen & Knob
      Mode PACV, PSIMV, VACV, VSIMV, Spont, Apnea Back-up ventilation
        O2Stream®, PRVC, Bi-Level, AutoVent® (APRV)
      Optional Nasal Cannular for O2Stream®
        Proximal(Adult/Neonate) Flow Sensor / SpO2 / ETCO2
2. Detalil Specifications
  2.1 Ventilator Data (Setting Parameters)
    Predicted Body weight   ~ 150kg
    Tidal Volume   5 to 2500 mL
    Inspiratory pressure   5 to 80 cmH2O
    Pressure support   0 to 60 cmH2O
    Respiratory rate   2 to 150 bpm
    Inspiratory time   0.2 to 9.9 seconds
    Pause time   0 to 2.0 seconds
    PEEP/CPAP  

0 to 45 cmH2O

    Enable inhalation trigger(En-sense) 10 to 80% of set tidal volume
    Exhalation trigger Sensitivity(Ex-trig) 10 to 99% of exhaltion time
    F-end (Flow end)   25 to 100% of peak flow
    Trigger sensitivity   Pressure : 0.5 to 20 cmH2O
        Flow : 0.5 to 20 lpm
    FiO2%   21 to 100%
    Sigh   OFF or Delivers one sigh breath every 30 / 60/ 90/ 120 breaths
        Sigh volume = Set tidal volume x 1.5
    Inspiratory Rising Time   0.1 to 0.5 sec
    Inspiratory Rising Time PSV   0.1 to 0.5 sec
    Inspiratory Pressure limit   10 to 80 cmH2O in PRVC and O2 Stream
    i-PAP (t-Bilevel)   3 to 45 cmH2O
    e-PAP (t-Bilevel)   0 to 42 cmH2O
    AutoVent®  (Minute Volume)   70 to 300%
    O2 Stream®  (Inspiratory flow rate)   5 to 40 lpm
    BTPS   OFF, Auto Dry / Auto Humid
    MASK ventilation (NIV)   ON / OFF
    Proximal Flow Sensor   ON / OFF
    High Frequency Ventilation Mode    
    Power (Single / Dual)   1 to 100%
    High Frequency (Single / Dual)   2 to 20Hz
    IMV Ex duration (Dual)   0.3 to 6.0 sec
    ExRate (Dual)   2 to 120 bpm
  2.2 Display Data
    Parameters   Setting parameters/patient status parameters/Alarm status/I:E ratio
    Graphic Waveform  

Pressure-Time, Flow-Time, Volume-Time

      Loops

Pressure-volume, Flow-volume, Pressure-Flow

    Special Measurement   P0.1 measurement, RSBI (Rapid Shallow Breathing Index), WOB
    Alarm   Ti/Ttot, Expiratory flow, Event log, Compliance
        O2/Air supply pressure Fail / Obstructed tube / Ventilator in-operation
        Circuit open
  2.3 Alarm Setting Parameters (MAIN)
    High tidal volume (Expiratory volume) OFF, 5 to 2500 ml
    Low tidal volume (Expiratory volume) 0 to 2500 ml
    High minute volume 0.1 to 50 lpm
    (Expiratory minute volume)  
    Low minute volume 0.0 to 49.9 lpm
    (Expiratory minute volume)  
    High respiration rate   3 to 180 bpm
    Low respiration rate   2 to 179 bpm
    High peak airway pressure   1 to 120 cmH20
    Low peak airway pressure   0 to 119 cmH20
    High O2%   OFF, 1 to 100%
    Low O2%   0 to 100%
    Airway leak   OFF, 50 to 500 ml
    Apnea   OFF, 10 to 60 sec.
    (Option, SPO2)        
    High SpO2   OFF, 52 to 99%
    Low SpO2   OFF, 51 to 99%
    High pulse rate (SpO2)   30 to 250 bpm
    Low pulse rate (SpO2)   25 to 245 bpm
    (Option, EtCO2)        
    High EtCO2   0.0 to 10.0 mmHg
    Low EtCO2   OFF, 0.0 to 9.9 mmHg
    High inspired CO2   0.0 to 10.0 mmHg
    Low inspired CO2   OFF, 0.0 to 9.9 mmHg
    High respiration rate (EtCO2)   3 to 150 bpm
    Low respiration rate (EtCO2)   2 to 149 bpm
  2.4 Key, Indicators        
    Nebulizer   10 to 180 minute or automatic
    Inspiration hold   Measures patient's lung compliance and resistance
        Elasticity, Time constant
    Expiration hold  

Measures auto-PEEP

    Manual inflation   Delivers one mandatory breath
    100% O2   Delivers 100% oxygen for 3 minutes
    Key lock   Key & LCD Touch Screen Lock
    Graph freeze        
    Alarm silence  

Mutes audible alarms for 2 minutes

    Alarm reset   Clears visual indicators and messages
  2.5 Electrical Requirements        
    Power   AC : 100-230 VAC, 50/60 Hz (Free voltage)
        DC : 12VDC 7A, Internal battery
            Provides Max. 180 minutes of normal operation
  2.6 Physical Characteristic        
    Weight   57 kg (main unit with LCD monitor w/ cart)
    Dimension Total 48cm W x 57cm D x 145cm H
      Pneumatic 33cm W x 25cm D x 40cm H
      LCD Monitor 31cm W x 4.8cm D x 28cm H
      Mobile Cart 48cm W x 57cm D x 77cm H
  2.7 Environmental Requirements        
    Operating temperature   10 to 40 ℃ (50 to 104 ℉)
    Storage temperature   -15 to 50 ℃ (5 to 122 ℉)
    Relative humidity   10 to 90% (storage)
  2.8 Pneumatic Requirements        
    Oxygen/Air inlet supply pressure   2.46 to 6.32 kgf/cm2 / 2.4 to 6.2 bar (35 to 90 psig)